Autoclave Engineers


s Corporate Home
Autoclave Engineers
Universal Reactor Controller - URC

Autoclave Engineers Universal Reactor Controller (URC) provides the temperature, pressure and mixer speed control required for the majority of our stirred and non-stirred reactor systems designed to operate in a laboratory environment.  This compact stand-alone controller is powerful enough to control large reactors with the addition of an external motor speed controller that is capable of handling larger DC & AC motors in excess of the URC’s internal ½ Horsepower capability.


Electrical equipment used to control any and all reactor functions.


Universal Ractor Controller - URC: Process Instrumentation
Fully integrated control, monitoring, and data handling of temperature, pressure, stirring speed, and other parameters for laboratory style reactors



FEATURES

  • PID control of electric heaters up to 20 Amps using internal circuitry. An optional 35 Amp remote power box is also available.

  • Speed control is for a Permanent Magnet DC motors from 1/8HP to 1/2HP.

  • Adjustable setpoint ramping and auto tuning of temperature

  • An Over-Temperature control function will shutdown the vessel’s electric heater if the skin temperature exceeds alarm setpoint.  An automatic reset function restarts the reactor when temperature cools below setpoint.

  • Optional communication package allows remote data accumulation and control. Works in conjunction with supervisory computer.

  • PID controlled Cooling Control Option controls external solenoid valve plumbed to vessel’s cooling coil.

  • Monitors and Controls Vessel Pressure, Inlet and Outlet Pressure Control is available using 2-way valves or electronically actuated metering valves.

  • 120/240 VAC, 50/60 Hz supply power capability


BENEFITS


The URC is capable of handling all sizes of AED reactors from the Mini-Reactor to the 4 Liter Vessels.

Heating, Cooling, Mixer Speed and Pressure Control PID Loops all have a push button, Auto-Tune feature that will configure each control loop for optimum process control.

The URC can be used for intrinsically safe operation when connected to an Autoclave Engineers Isolator Package.

 

 





Match:
any search words
all search words


Results per page:

Autoclave Engineers
8355 Hessinger Drive
Erie, Pennsylvania 16509 USA
Phone: 814.860.5700
Fax: 814.860.5811

Contact Us

Valves, Fittings, and Tubing Looking for Valves, Fittings and Tubing?

fg Click here to view our sister division...
Autoclave Engineers
Fluid Components

Bookmark and Share



2012©All Rights Reserved • Autoclave Engineers • Erie, PA USA
Division of Snap-tite Inc.